Advantages
Increased laboratory efficiency
A constant, uninterrupted gas supply of guaranteed purity eliminates interruptions of analyses to change cylinders and reduces the amount of instrument re-calibrations required.
Improved economy
Pure nitrogen gas produced as standard
Improved safety
Nitrogen produced at low pressure and ambient temperature removes the need for high pressure cylinders
Security of supply
Integral oil free air compressor as an option guarantees continuous gas supply, independent of in house compressed air supply
Simple installation
Gas generators can be installed in the laboratory, on or under a bench, eliminating the need for long gas lines from cylinders secured elsewhere
Characteristics
- Fully regenerative PSA technology; reduced risk of gas contamination and phthalate free
- Various flow rate
- Options :
- Integral oil free air compressor
- Oxygen analyser
- Catalyst module for hydrocarbons level < 0.05ppm - Quiet thanks to the Soundproofed compressor boxand anti-vibration features
- Auto start
- Audible and Alarm display with help menu and history log
- Outlet flow indicator
- Visual maintenance indication with history log
- Energy saving Mode: Enables the compressor to switch off when nitrogen supply is not required
- Remote acess to screen using internet via USB port. To learn more, click here
Product Features
- Generates Nitrogen on demand from external compressed air
- Designed to run 24 hours a day
- Eliminates dangerous Nitrogen cylinders from the laboratory
- The ZN2 series is equipped with a Catalyst oven to remove Hydrocarbons (as methane)
- Smart control of the flow of inlet air, it is function of the real flow of nitrogen used by the application
- Minimum maintenance with an annual filter change
- USB/RS232 ports to remote the system anywhere in the world
- Low power, low noise, low maintenance
- Ultra fast start-up time
- Fast purity (10 ppm in less than 3 hours)
Main Applications
- Produces a supply of high purity Nitrogen 99.9995% for GC application
- TOC applications including combustion, UV persulfate and wet oxidation sampling techniques
- make-up and carrier gas applications including ECD
Technical Specifications
Model: | N2.600.HP | N2.600.HP.D | N2.600.HP.D | N2.1000.HP | N2.1000.HP.D |
N2 outlet | |||||
Flow rate (max) | 600 cc/min | 1000 cc/min | |||
Outlet pressure (min) | 1.0 bars | ||||
Outlet pressure (max) | Inlet pressure – 2.0 bar | ||||
Nitrogen purity | > 99.9995% – (<5ppm O2 @ 7 Bar inlet pressure) | ||||
O2 content | – | – | < 5 ppm | – | – |
Hydrocarbon content (CH4 free) | – | – | < 0.05 ppm | – | – |
Air intel | |||||
Air inlet requirement | Particulate: <1 micron Dewpoint: < -40°C (-40°F) Oil: <0.01 mg/m³ |
||||
Flow rate (max) | 12 l/min | 15 l/min | |||
Min supply pressure | 7 bars | ||||
Max supply pressure | 10 bars | ||||
Min temperature | 1°C (34°F)/td> | ||||
Max temperature | 35°C (95°F)/td> | ||||
Communication | |||||
LCD with touch screen | No | Yes | Yes | No | No |
RS232 | X | ||||
RS485 | X | ||||
USB | X | ||||
General data | |||||
Supply rating | 100-240Vac 50/60Hz – 2A | ||||
Connection type | IEC320-C13 | ||||
Installation power (max) | 240W (280VA) | ||||
Fuse rating | 4A-240Vac (5x20mm) | ||||
Net weight | <25 kg | ||||
Connections | |||||
Outlet port | 1/4″ female | ||||
Inlet port | 1/4″ female | ||||
Operating/storage conditions | |||||
Temperature | 5-35°C (41-95°F) | ||||
Humidity (max, non condensing) | 80% at 25°C (77°F) | ||||
Noise | <25dB(A) | ||||
IP rating | IP20 | ||||
Pollution degree rating | 2 (with no aromatic compounds) | ||||
Altitude | < 200 |