N2 Generator Desk
N2.600.HP.D / ZN2.600.HP.D / N2.1000.HP.D

The MicroPROGEL High Purity (HP) Nitrogen Generators are engineered to transform standard compressed air in to a safe regulated supply of 99.9995% pure nitrogen. Typical applications include GC make up gas and carrier gas, including ECD (Electron Capture Detector), solvent evaporation, DSC (Differential Scanning Calorimeter) and virtually any analytical instrument that requires a small flow of ultra high-purity nitrogen.

 Download N2 Generator Desk Tech Sheet

Advantages

Increased laboratory efficiency

A constant, uninterrupted gas supply of guaranteed purity eliminates interruptions of analyses to change cylinders and reduces the amount of instrument re-calibrations required.

Improved economy

Pure nitrogen gas produced as standard

Improved safety

Nitrogen produced at low pressure and ambient temperature removes the need for high pressure cylinders

Security of supply

Integral oil free air compressor as an option guarantees continuous gas supply, independent of in house compressed air supply

Simple installation

Gas generators can be installed in the laboratory, on or under a bench, eliminating the need for long gas lines from cylinders secured elsewhere

Characteristics

  • Fully regenerative PSA technology; reduced risk of gas contamination and phthalate free
  • Various flow rate
  • Options : 
    - Integral oil free air compressor 
    - Oxygen analyser 
    - Catalyst module for hydrocarbons level < 0.05ppm
  • Quiet thanks to the Soundproofed compressor boxand anti-vibration features
  • Auto start
  • Audible and Alarm display with help menu and history log
  • Outlet flow indicator
  • Visual maintenance indication with history log
  • Energy saving Mode: Enables the compressor to switch off when nitrogen supply is not required
  • Remote acess to screen using internet via USB port. To learn more, click here

Product Features

  • Generates Nitrogen on demand from external compressed air
  • Designed to run 24 hours a day
  • Eliminates dangerous Nitrogen cylinders from the laboratory
  • The ZN2 series is equipped with a Catalyst oven to remove Hydrocarbons (as methane)
  • Smart control of the flow of inlet air, it is function of the real flow of nitrogen used by the application
  • Minimum maintenance with an annual filter change
  • USB/RS232 ports to remote the system anywhere in the world
  • Low power, low noise, low maintenance
  • Ultra fast start-up time
  • Fast purity (10 ppm in less than 3 hours)

Main Applications

  • Produces a supply of high purity Nitrogen 99.9995% for GC application
  • TOC applications including combustion, UV persulfate and wet oxidation sampling techniques
  • make-up and carrier gas applications including ECD

Technical Specifications

Model: N2.600.HP N2.600.HP.D N2.600.HP.D N2.1000.HP N2.1000.HP.D
N2 outlet
Flow rate (max) 600 cc/min 1000 cc/min
Outlet pressure (min) 1.0 bars
Outlet pressure (max) Inlet pressure – 2.0 bar
Nitrogen purity > 99.9995% – (<5ppm O2 @ 7 Bar inlet pressure)
O2 content < 5 ppm
Hydrocarbon content (CH4 free) < 0.05 ppm
Air intel
Air inlet requirement Particulate: <1 micron
Dewpoint: < -40°C (-40°F)
Oil: <0.01 mg/m³
Flow rate (max) 12 l/min 15 l/min
Min supply pressure 7 bars
Max supply pressure 10 bars
Min temperature 1°C (34°F)/td>
Max temperature 35°C (95°F)/td>
Communication
LCD with touch screen No Yes Yes No No
RS232 X
RS485 X
USB X
General data
Supply rating 100-240Vac 50/60Hz – 2A
Connection type IEC320-C13
Installation power (max) 240W (280VA)
Fuse rating 4A-240Vac (5x20mm)
Net weight <25 kg
Connections
Outlet port 1/4″ female
Inlet port 1/4″ female
Operating/storage conditions
Temperature 5-35°C (41-95°F)
Humidity (max, non condensing) 80% at 25°C (77°F)
Noise <25dB(A)
IP rating IP20
Pollution degree rating 2 (with no aromatic compounds)
Altitude < 200