Aurora 1030W TOC Analyzer

Significantly expand sample range and increase throughput with the Aurora 1030W TOC Analyzer, an instrument specifically designed for analyzing organic contamination levels in aqueous samples by the heated persulfate wet oxidation technique.

 

The Aurora 1030W TOC Analyzer uses the proven heated persulfate wet oxidation technique to analyze organic contamination levels in liquid samples. The instrument can be programmed and calibrated to analyze samples containing organic carbon at levels as low as 10 ppb and as high as 30,000 ppm. In addition to its wide range, the Aurora 1030W has an extremely high daily throughput. Up to 300 samples can be analyzed in a 24-hour period (dependent on method and application).

The reaction chamber of the Aurora 1030W is thoroughly rinsed between analyses to eliminate residue from the previous sample. This ensures the Aurora 1030W maintains the low system background necessary for high sensitivity TOC measurements. This contrasts with combustion TOC analyzers in which residual salts accumulate on catalyst surfaces, degrading oxidation efficiency and causing higher blanks and background contamination.

Tested & Approved

The heated persulfate oxidation technique has been approved and adopted in numerous regulatory compliance methods, standards, and norms for water quality testing, including: Standard Method 5310C, USEPA 415.3, USP <643> / EU 2.2.44, ASTM D 4779 and 4839, USEPA-DBPR, USEPA-SPCC, ISO 8245, and EN 1484.

Applications

The Aurora 1030W TOC Analyzer is ideal for the following applications: drinking water, pharmaceutical cleaning validation, municipal wastewater, ground water/surface water, process water, boiler feed water and condensate, metal plating solutions, and ultrapure water.

Aurora 1030W Features:

  • Wide sample range from 10 ppb to 30,000 ppm
  • High sample efficiency – Up to 300 samples daily
  • Rinse function eliminates residue from previous sample preventing blanks and background contamination
  • Heated persulfate oxidation technique is compliant with numerous regulatory methods, standards, and norms
  • Operates with an intuitive color touchscreen display with Windows® CE-based software
  • Options available to process more samples and broaden scope of sample type

TOC WEBINAR SERIES:
View webinar recordings now!
Play-Webinar-button.png

 

Wet Oxidation Mode
 
Operating Principle Heated sodium persulfate oxidation
Measurement Range 10 ppb C - 30,000 ppm C (multiple calibration ranges or dilution required)
Instrument Detection LImit (IDL) 2 ppb C
Sample Injection Volume 10µL – 10mL
Method TC Acid and persulfate reaction
Method TIC Acidification with Phosphoric acid and sparging
Method TOC NPOC by heated persulfate oxidation or TC - TIC
Reagents Required Sodium persulfate, 5% phosphoric acid, rinsewater
Gas Supply N2 (99.998%), zero-grade air, or O2 (99.998%)

 

General Instrument Specifications  
Measurement Technique Non-dispersive infrared (NDIR) detector
Operator Interface Color LCD touchscreen display with Windows® CE-based software
Basic Software Single instrument operation with data transfer to PC
Optional ATOC Software Network LAN/LIMS operation, data management, custom reports, 21 CFR 11 compliance
Autosampler 88 position rotary autosampler designed to fit directly underneath Aurora 1030D analyzer
Certification CE, EMC: EN61326 / Safety: IEC 61010-11 2001
Power Supply Variable voltage, 100-240VAC, 50/60Hz, 950W
Dimensions – Aurora 1030D + 1088 Autosampler 26.75 in. H x 19.5 in. W x 23 in. D.
Weight – Aurora 1030D + 1088 Autosampler 34.5 kg (76 lbs.)